TRIVAC D 8 BPumping speed 8.5 m³/hUltimate pressure 2 × 10-3 mbarTRIVAC D 65 BCS 3Ph 200-240/380-400 V 50 Hz / 200-240/400-480 V 60 Hz
The Leybold TRIVAC D 65 BCS is the corrosion-protected version of the largest TRIVAC B: a two-stage, oil-sealed rotary vane vacuum pump with a pumping speed of 65 m³/h (50 Hz) and DN 40 ISO-KF connections for corrosive or aggressive process gases. As a PFPE version it is prepared for LEYBONOL LVO 400. Ehrler & Beck offers the Leybold TRIVAC D 65 BCS in two variants, as well as the inert gas system, oil filters and spare parts.
Request a quote for this product Read more
- Pumping speed 65 m³/h 50 Hz 78 m³/h 60 Hz
- Ultimate pressure 2 × 10-3 mbar
- Rated motor power 2.2 kW 50 Hz 2.2 kW 60 Hz
| Variant | Part number |
|---|---|
| 3Ph 200-240/380-400 V 50 Hz / 200-240/400-480 V 60 Hz (selected) | 11398 |
| 3Ph 50/60 Hz, BCS-PFPE (for LEYBONOL LVO 400) | 11399 |
TRIVAC D 65 BCS – corrosion-protected version with 65 m³/h
The Leybold TRIVAC D 65 BCS combines the pumping speed of the TRIVAC D 65 B (65 / 78 m³/h at 50 / 60 Hz, ultimate total pressure ≤ 2 × 10⁻³ mbar, DN 40 ISO-KF) with the corrosion-protected equipment of the BCS series. It is intended for pumping corrosive or aggressive media in the chemical industry and semiconductor manufacturing, for research and production, and as a backing pump in pump systems.Design and function
According to Leybold, the TRIVAC BCS is of corrosion-protected design: yellow metals have been dispensed with; grey cast iron, surface-treated aluminium, steel and stainless steel are used. Added to this are a double shaft seal, a built-in temperature switch for temperature monitoring, the built-in oil pump for continuous operation up to 1000 mbar and the oil-pressure-controlled anti-suckback valve. Together with the chemical oil filter with safety shut-off valve CFS 40-65, the exhaust filter with oil return ARS 40-65, the inert gas system IGS 40-65 and the limit switch system LSS 40-65, the pump forms the TRIVAC SYSTEM. Water vapour tolerance 40 mbar, water vapour capacity 1925 g/h (50 Hz) / 2310 g/h (60 Hz), oil filling 2.0 / 3.3 l, noise level 57 / 59 dB(A) (without / with gas ballast).PFPE version
The BCS-PFPE variant is prepared for operation with perfluoropolyether LEYBONOL LVO 400 and is supplied without an oil filling. According to Leybold, PFPE is practically inert to chemical and oxidising influences, does not polymerise under high-energy radiation and is thermally highly stable (decomposition only above 290 °C). Leybold recommends always installing a chemical oil filter CF or CFS. Ultimate partial pressure without gas ballast: < 8 × 10⁻⁴ mbar.Variants of the TRIVAC D 65 BCS
The TRIVAC D 65 BCS is available from Ehrler & Beck in two variants (part numbers 11398, 11399). The rated motor power is 2.2 kW (50 Hz), the weight ranges from 84.5 to 87.9 kg. Voltage, phase and all further data of the individual variants can be found in the variant selection and under Technical data.Features and benefits
- Corrosion-protected: No yellow metals; grey cast iron, surface-treated aluminium, steel and stainless steel
- Double shaft seal and built-in temperature switch for temperature monitoring
- TRIVAC SYSTEM: Matched to the chemical oil filter CFS, exhaust filter ARS, inert gas system IGS and limit switch system LSS
- PFPE version: Prepared for LEYBONOL LVO 400
- Continuous operation up to 1000 mbar: Built-in oil pump, pressure-lubricated plain bearings
- Low oil backstreaming: Oil-pressure-controlled anti-suckback valve, independent of gas ballast operation
| Pumping speed 50Hz | 65 |
|---|---|
| Pumping speed 60Hz | 78 |
| Ultimate pressure | 2 × 10-3 |
| Rated motor power 50Hz | 2.2 |
| Rated motor power 60Hz | 2.2 |
| Motor efficiency class | IE3 |
| Weight | 84.5 |
| Length | 79.7 |
| Width | 25.5 |
| Height | 35.7 |
| Inlet connection flange | DN 40 ISO-KF (NW40) |
| Outlet connection flange | DN 40 ISO-KF (NW40) |
| Noise level | 57 dB(A) |
| Cooling method | Air cooling |
| Oil type | LEYBONOL LVO 100, LEYBONOL LVO 400 |
Additional technical data
| Technical data | Value |
|---|---|
| Nominal pumping speed | 75 m³/h (50 Hz) / 90 m³/h (60 Hz) |
| Ultimate partial pressure without gas ballast | < 1 × 10⁻⁴ mbar |
| Ultimate total pressure without gas ballast | ≤ 2 × 10⁻³ mbar |
| Ultimate total pressure with gas ballast | < 5 × 10⁻³ mbar |
| Water vapour tolerance | 40 mbar |
| Water vapour capacity | 1925 g/h (50 Hz) / 2310 g/h (60 Hz) |
| Oil filling min. / max. | 2.0 / 3.3 l |
| Noise level | 57 / 59 dB(A) |
| Nominal speed 50 / 60 Hz | 1460 / 1760 min⁻¹ |
| Mains voltage | 3Ph 200 – 240/380 – 400V 50 Hz / 3Ph 200 − 240V/ 400 − 480V 60 Hz |
| Mains phases | 3-Ph |
| Power consumption at ultimate pressure | 1.32 kW |
| IP protection class | IP 55 |
| Leak rate | 1 × 10⁻⁴ Pa·m³/s |
| Permissible ambient temperature | +12 to +40 °C |
| Operating altitude, max. | 1000 m |
| Motor certification | CE, UL (61010-1 3rd Edition) |
| Materials of construction | Steel, cast iron, aluminium, FPM (FKM), glass, filter material (polymers, paper), epoxy resin and glass fibre, copper |
| Product name | Part number |
|---|---|
| Pump module TRIVAC D 65 BCS | E20039760 |
| Major maintenance kit (without oil) TRIVAC D 65 BCS | EK110002642 |
| Minor maintenance kit D 40 B / D 65 B / BCS | EK110002624 |
| Radial shaft seal kit D 40/65 BCS | EK110002643 |
| Seal kit S/D 40-65 BCS | 19732 |
| Spare parts kit TRIVAC D 40/65 BCS | EK110002651 |
| Major maintenance kit (without oil) TRIVAC D 65 BCS-PFPE | EK110002638 |
| Seal kit S/D 40-65 BCS-PFPE | 19742 |
| Product name | Part number |
|---|---|
| Roots pump adapter for RUVAC WA(U)/WS(U) 251/501 | 16830 |
| Inert gas system IGS 40-65 UL | 16168V |
| Chemical oil filter with safety shut-off valve CFS 40-65 | 10177 |
| Exchange filter WF Alu-Part for CFS (2 required) | 18999 |
| Exhaust filter with oil return ARS 40-65 | 18957 |
| Replacement filter element FE 40-65 for ARS 40-65 | 18973 |
| Condensate trap AK 40-65 DN 40 KF | 18816 |
| LEYBONOL LVO 400 (PFPE), 1 litre | L40001 |














