Edwards E2M0.7Pumping speed 0.75 m³/hUltimate pressure 3 × 10-3 mbarEdwards E2M175 HC: 3Ph 380/400 V 50 Hz, 230/460 V 60 Hz, IE3 (Europe/USA)
The Edwards E2M175 is a robust two-stage oil-sealed rotary vane vacuum pump with a pumping speed of 160 m³/h at 50 Hz (196 m³/h at 60 Hz) and an ultimate pressure of 3.3 × 10⁻³ mbar without gas ballast. The pump is water-cooled. It is available as a standard version and as an azide-compatible version. Ehrler & Beck advises on selection, accessories and service.
Request a quote for this product Read more
- Pumping speed 160 m³/h 50 Hz 196 m³/h 60 Hz
- Ultimate pressure 3.3 × 10-3 mbar
- Rated motor power 5.5 kW 50 Hz 6.5 kW 60 Hz
| Variant | Part number |
|---|---|
| HC: 3Ph 380/400 V 50 Hz, 230/460 V 60 Hz, IE3 (Europe/USA) (selected) | A36604940 |
| HC: 3Ph 200 V 50 Hz, 200/380 V 60 Hz, IE3 (Asia) | A36604934 |
| HC: without motor, free shaft end with IEC flange | A36601000 |
| AZ: 3Ph 380/400 V 50 Hz, 230/460 V 60 Hz, IE3 (Europe/USA) | A36607940 |
| AZ: 3Ph 200 V 50 Hz, 200/380 V 60 Hz, IE3 (Asia) | A36607934 |
Edwards E2M175 – two-stage rotary vane vacuum pump for industry
The E2M175 belongs to Edwards' E2M series of robust two-stage rotary vane pumps (40 to 275 m³/h). It delivers 160 m³/h at 50 Hz and 196 m³/h at 60 Hz and reaches an ultimate pressure of 3.3 × 10⁻³ mbar without gas ballast, 0.013 mbar with gas ballast. The motor power is 5.5 kW at 50 Hz and 6.5 kW at 60 Hz (Europe/USA version). The sound pressure level is 75 dB(A); with motor the pump weighs a maximum of 204 kg and measures 1039 × 415 × 564 mm (L × W × H). The E2M175 is water-cooled. Edwards recommends at least 80 l/h of cooling water at 20 °C; with the optional thermostatic valve the pump reaches its operating temperature faster and tolerates more water vapour.Features of the E2M series
- Robust design with performance proven over many years in demanding markets
- Good sealing of the vacuum stage for a low and stable ultimate pressure
- Adjustable gas ballast with built-in dust filter and purge gas connection
- Forced lubrication for safe operation at any inlet pressure; oil pressure can be checked with a pressure gauge
- Built-in oil filter; external oil filtration possible
- Inlet isolation valve against oil suck-back; optional oil mist filters at the outlet
Typical applications
For the E2M pumps Edwards names, among others, coating, freeze drying in pharmaceuticals (AZ version for azide compatibility), plasma treatment (FX version for oxygen-rich plasmas), heat pipes, heat treatment and the chemical process industry.Versions of the Edwards E2M175
The E2M175 is available in 5 versions (part numbers A36604940, A36604934, A36601000, A36607940, A36607934). HC is the standard pump for hydrocarbon oil; on the AZ version copper and brass parts are replaced by suitable materials such as stainless steel (azide-compatible). The motor versions are available for Europe/USA and for Asia; the versions without motor have a free shaft end with IEC flange. The market details come from the Edwards ordering documents.Advice and service
Ehrler & Beck supports you in selecting the right version, supplies oil, accessories and spare parts and carries out maintenance and repair.| Pumping speed 50Hz | 160 |
|---|---|
| Pumping speed 60Hz | 196 |
| Ultimate pressure | 3.3 × 10-3 |
| Rated motor power 50Hz | 5.5 |
| Rated motor power 60Hz | 6.5 |
| Motor efficiency class | IE3 |
| Weight | 204 |
| Length | 103.9 |
| Width | 41.5 |
| Height | 56.4 |
| Inlet connection flange | DN 63 ISO-F |
| Outlet connection flange | DN 40 ISO-F |
| Noise level | 75 dB(A) |
| Cooling method | Water cooling |
| Oil type | Ultragrade 70 |
Additional technical data
| Technical data | Value |
|---|---|
| Design | Two-stage oil-sealed rotary vane vacuum pump, water-cooled |
| Version | HC: 3Ph 380/400 V 50 Hz, 230/460 V 60 Hz, IE3 (Europe/USA) |
| Max. pumping speed (Pneurop) 50 Hz / 60 Hz | 160 / 196 m³/h |
| Max. pumping capacity (displacement) 50 Hz / 60 Hz | 178 / 214 m³/h |
| Ultimate pressure without gas ballast | 3.3 × 10⁻³ mbar |
| Ultimate pressure with gas ballast | 0.013 mbar |
| Max. water vapour inlet pressure | 4 mbar (20 mbar with thermostatic valve) |
| Max. water vapour pumping capacity | 2.4 kg/h |
| Max. gas ballast flow | 3 m³/h |
| Max. permissible outlet pressure | 0.5 bar gauge (1.5 bar absolute) |
| Motor power 50 Hz / 60 Hz | 5.5 / 6.5 kW |
| Motor efficiency class | IE3 |
| Motor speed 50 Hz / 60 Hz | 1440 / 1720 min⁻¹ |
| Sound pressure level (1 m, ultimate pressure) | 75 dB(A) |
| Weight (with motor, without oil) | max. 204 kg |
| Dimensions L × W × H | 1039 × 415 × 564 mm |
| Suction connection | ISO63 (flange with trapped O-ring) |
| Outlet | ISO40 (flange with central 1½" BSPP thread) |
| Cooling | Water cooling, connections 2 × 3/8" BSPP/T, max. 3 bar, recommended minimum cooling water flow 80 l/h at 20 °C |
| Oil | ULTRAGRADE 70 |
| Oil filling (min.–max.) | 16–25 l |
| Operating temperature (ambient) | 12 to 40 °C |
| Type of protection | IP55 |
| Product name | Part number |
|---|---|
| Oil ULTRAGRADE 70 (HC and AZ versions) | H11028016 |
| Product name | Part number |
|---|---|
| Inlet dust filter ITF300 | A44203000 |
| High-capacity inlet dust filter ITM300 | A44303000 |
| Inlet catchpot ITO300 | A44103000 |
| Inlet chemical trap ITC300 | A44403000 |
| Outlet oil mist filter MF300 (clean applications) | A46204000 |
| Oil return kit for outlet oil mist filter | A50005000 |
| Outlet catchpot CP300 | A46104000 |
| Vibration isolators | A24801406 |
| Thermostatic valve for the cooling water (brochure: TCU300) | A50001000 |