Edwards nXL110iD Enhanced 1Ph 200–240 V, 50/60 Hz, main inlet NW40, secondary inlet NW25, outlet NW25

Part number: A77072320 Series: nXLi series Manufacturer: Edwards Technology: Roots vacuum pumps Industries: Research, laboratory and analytics Applications: Mass spectrometry and analytics

The Edwards nXL110iD Enhanced is an oil-free, air-cooled dry pump (multi-stage Roots pump) with a second inlet port and a pumping speed of 98 m³/h at the main inlet; the ultimate pressure is below 0.07 mbar. Ehrler & Beck advises on selection and service.

Request a quote for this product Read more

  • Pumping speed 98 m³/h 50 Hz 98 m³/h 60 Hz
  • Ultimate pressure 0.07 mbar
Available variants
Variant Part number
1Ph 200–240 V, 50/60 Hz, main inlet NW40, secondary inlet NW25, outlet NW25 (selected) A77072320
1Ph 200–240 V, 50/60 Hz, main inlet NW50, secondary inlet NW25, outlet NW25 A77072420

Edwards nXL110iD Enhanced – dry vacuum pump with two inlets

The nXLi Enhanced pumps combine the proven multi-stage Roots technology with a further developed drive. The nXL110iD Enhanced reaches 98 m³/h at the main inlet and 78 m³/h at the secondary inlet and an ultimate pressure below 0.07 mbar. The second inlet allows two vacuum areas to be served with one pump.

Features

  • Transient Boost Mode: higher performance for a defined period, for example during the first pump-down or with a briefly higher gas load; according to Edwards gas loads up to 60 slm nitrogen and chambers up to 300 litres
  • Single-phase supply 200–240 V ±10 %, 50/60 Hz, for worldwide use
  • Air-cooled with controlled cooling that adapts the cooling capacity and keeps the noise low
  • Control panel as on the other Edwards dry pumps; remote and local operation, USB port for monitoring and diagnostics with the nST2 software
  • Oil-free and dust-free pumping chamber for analytics and research

Versions of the Edwards nXL110iD Enhanced

The versions differ only in the flange of the main inlet (NW40 or NW50).

Advice and service

Ehrler & Beck supports you in selection and supplies accessories and service.